Care & Handling
To Our Customers:
in reference to why staining, pitting, or cracking can develop on stainless steel instruments; there are some basic tasks for the care and handling of surgical instruments which will extend the life of the device, eliminate staining or corrosion and lower usage-costs.
Care & Handling Instruments
- Instruments should be rinsed as soon as possible after use to remove blood, serum and saline. Do not use a sponge basin as any residue from normal saline solution can cause corrosion to develop over a period of time.
- Instruments should be thoroughly dried after cleaning and before wrapping or storing, as moisture left in the box lock or serrations can cause rust or corrosion to develop.
- Do not leave instruments in a cold sterilization soak solution for extended periods as these solutions tend to be corrosive (after cold sterilization, rinse the instruments in distilled or demineralized water and dry thoroughly).
- If using cold sterilization technique, change the solution according to directions, as prolonged use will cause the solution to become corrosive.
- Do not use caustic cleaners or any other cleaner except those specified for cleaning surgical instruments.
- Do not clean instruments with steel wool or scouring powder, or use high PH detergents. If cleaning by hand, use a specific-instruments cleaner or a neutral-detergent soap and a soft brush (rinse with demineralized or distilled water).
- Check your autoclave. iron, sodium, calcium, magnesium or copper in your water can cause spotting, staining or corrosion to occur. In hard water areas, clean the lime deposits from steam lines periodically or use distilled or demineralized water for your autoclave. Install a steam filter, if possible. It will remove most of dirt, rust and pipe scale.
- Make sure your Muslim wraps are rinsed sufficiently (6 to 8 times) by the laundry to remove all chlorine bleach as any residue can cause staining or corrosion during autoclave.
- Do not leave instruments in the following solutions for extended periods as corrosion can result:
- Alumium, Barium, Calcium, Ferrous or Stannous Chloride solutions.
- Phenol, Lysol or Lodine.
- Dakin's Solution or Zephirin.
- Any acid, Mercury or Potassium solution.
- Do not sterilize stainless steel instruments with chrome-plated instruments, if possible, as any break in the surface of the chrome plate will allow an electrolytic action to develop between the dissimilar metals, causing pitting and/or a rusty appearance to develop.
- Preheat instruments in autoclave before allowing steam to enter, if possible, to prevent an excessive amount of condensation forming on the surface of the instruments.
- Do not open the autoclave too quickly after the sterilization cycle as excessive condensation will develop on the surface of the instruments (crack the clave to allow drying before opening fully).
By following the above rules and using instrument cleaners and lubricant milk such as our Ami-Clean or Ami-Tec products, the instruments should remain rust, spot and stain-free.
We hope this will answer any questions regarding this topic.When proper care is taken these stainless steel instruments can last for many years.